The scheme shows an Aquinity P35/70 Life Science with integrated TOC-monitoring.
To save valuable benchtop space the system can be ordered as BI configuration. The system will be installed underneath the work bench and only the display and dispenser are mounted on the wall.
To save valuable benchtop space the system can be ordered as wall mounted configuration. the whole system can be mounted on the wall with inclined display.
Compact pump is required when feeding laboratory equipment such as autoclaves or laboratory dish washers.
The TOC monitoring during production and intermittent measurements during non-use periods allows to check the organic content in water continuously. The TOC value is measured between 1 and 999 ppb and can be shown in the display to monitor organic impurities.
Our dispenser allows the volumetric controlled dispensing of water with an increment of 0.1 L and a tap volume from 0.1 to 99 L. The system prevents overflow of containers and allows to dispense water without supervising.
Validation documents for water purification systems.
To be integrated into an ultra pure water system with a max, flow rate of 2 L/min. Point of use dispense arm with integrated control unit. Two joint to move it horizontally and vertically allowing to fill containers up to 55 cm height. Volumetric dosing of ultra pure water allows filling of containers without the attention of user, dosing eange 0.1 – 99 L, variable input and storage of values, after volume has been dispensed system goes back to standby.
Technical Specifications | |
---|---|
pure water + ultrapure water quality | < 2 µS/cm; Type II + 0.055 µS/cm; Type I |
resistivity | > 0.5 MegaOhm; Type II + 18.2 MegaOhm/cm; Type I |
total organic carbon (TOC) | < 10 ppb Reagent, < 5 ppb Life Science, < 3 ppb Analytical |
dispensing flow rate | 2.0 L/min (1.5 L/min Life Science) |
production rate | 10 L/h, optionally 20 L/h |
bacteria | < 1 cfu/mL |
particulate | > 0.2 µm less than 1 particulate/mL |
pyrogen (endotoxins) | < 0.001 EU/mL* |
RNAse | < 1 pg/mL* |
DNAse | < 5 pg/mL* |
dimensions, weight, power | 504 x 680 x 535 mm, 16-20 kg, 110 - 230 V |
*Life Science |
Feed Water Requirements | |
---|---|
type of feed water | tap water |
feed water conductivity | < 1400 µS/cm |
inlet pressure | 1.5 to 6 bar |
free chlorine | < 0.1 mg/L |
Silt Density Inded (SDI) | < 3 |
pH | 3 to 9 |
temperature | 5 to 25 °C |
The "Aquinity² P35/70" is a laboratory water system that provides its user with ultrapure water for a wide variety of applications in the field of analytics and laboratory applications. The designations "35" and "70" stand for an external tank with a capacity of 35 and 70 liters respectively.
The Aquinity² P35/70 systems produce pure and ultrapure water from drinking water. This makes the Aquinity² P35/70 the ideal system for a laboratory that pursues different types of applications that also have different requirements for the purity content of the laboratory water.
The compact design of the Aquinity²35/70 takes up little valuable laboratory space and its simple and intuitive operation makes it easy to integrate into everyday laboratory work.
Customers have the option of selecting a system with or UV lamp, depending on the area of application. If work is carried out in the life science sector, we also offer a configuration with ultrafiltration module. This eliminates the need for frequent changes of biological final filters and reduces investment costs.
Due to the simple and clear design of the software, the user can quickly operate the device through intuitive handling. The software informs the user when limit values have been exceeded or when consumables such as cartridges or UV lamps need to be replaced. A wizard guides the user through the individual steps.
In addition to the hoses for inlet and outlet, the scope of delivery also includes an adapter that makes it easy for them to connect the system to a drinking water connection.
Depending on the configuration, the system has a 35 or 70 liter tank as standard. Should a higher capacity be required, several tanks can also be connected in series.
By using an external tank, Type II water can be tapped through a separate tap. The tank also has a rear outlet so that external systems such as analyzers can be fed directly. A booster pump can also be offered on request to pump the water.
The installation location can be designed very flexibly. Depending on the available work surface, the system can be integrated directly on the equipment, on the wall or under the lab bench. The display, which can be installed flexibly, also ensures convenient operation.
We would like to improve the contact to our customers & partners and send you information and updates about our company and our products
membraPure endeavours as far as reasonably possible to provide correct and complete information on this website. However, membraPure does not guarantee that the information provided is up-to-date, correct and complete. The same applies to the content of external websites that this site refers to directly or indirectly via hyperlinks, programmes, files, contents or hyperlinks to other websites are provided without any expressed or implied warranty.
Contact Us
membraPure reserves the right to alter the information provided on this site without prior notice.
membraPure GmbH, Wolfgang-Kuentscher Str. 14, 16761 Hennigsdorf