Ion Chromatograph logo membraPure

Determination of DEA (Diethanolamine) in Refinery Boiler Water Systems

The Determination of DEA (Diethanolamine) in Refinery Boiler Water Systems Using the IONUS Ion Chromatograph

The process involves:

IC SystemIONUS Ion Chromatograph
ColumnRepromer Cat, 7 µm, 4 mm x 250 mm
Eluent3 mM HNO3 / 2.6 mM 18-Crown-6
Flow Rate 900 µL/min
Temperature40 °C
Injection20 µL
DetectionConductivity, 1 mS/cm
AnalyteConcentration
Sodium2.0
DEA2.9
Ammonium4.95
Figure 1: Ion Chromatograph IONUS
SampleNaDEANH4
168-10.060.4523.1
168-2Nd0.7115.2
168-30.900.079.98
172-10.010.6123.8
172-20.020.8014.8
172-30.900.0910.6
QC1.992.894.96
Std Values2.002.904.95

Fig. 1: Chromatogram of a multi-ion standard including Lithium, Sodium, DEA, TEA, MEA, Ammonium, Magnesium, and Calcium.
Fig. 2: Chromatograms of three different Refinery plant samples (168-1,2,3), displaying varying levels of sodium and DEA (Diethanolamine).

Complete your name and email to receive the application note.

Analysis of the amino acid composition of cell culture media
Get download link